Brite Technologies Ultimate Laboratory Solutions...

Photometer 5010 Made in Germany

TECHNICAL SPECIFICATIONS

 

Type:
Semi-automatic, single-beam filter photometer
Light Source:
Halogen lamp - 12 V, 20 W
Wavelength: 340 nm - 800 nm
Wavelength Selection: Automatic via 8-position filter wheel:
5 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm.
3 positions for optional filters of choice
Photometric Range: 0-2.5 Bel
Cuvette System: Microflow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control:

Internal Peltier elements, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C
Equilibration time for aspirated reaction mixture to reach 37 °C from ambient temperature: 15 sec

Aspiration System: Built-in peristaltic pump driven by stepper motor, programmable aspiration volume controlled by infrared light barrier
Shipping Volume: Minimum 250 µl
Operator Interface: Membrane keyboard with 20 keys for direct functions and numeric entries
Data Presentation: LCD display: lighted, 2 lines of 16 characters each, easy to read and offering simple user guidance
Integrated Printer: Dot-matrix printer, 24 characters per line
Memory: General operating software contained in FLASH MEMORY can be updated by PC (data file downloaded from Internet, floppy disk or CD ROM)
Languages: English and German
Capacity for 250 pre-programmed testvariants
Up to 230 tests can be stored in RAM. Import of data by keyboard or PC
Up to 50 non-linear calibration curves with max 20 sets of points can be stored
Signal Ports: 1 serial and 1 parallel port for connection to an external printer and/or computer (bidirectional interface)
Measurement Procedures: · Absorbance
· Endpoints with standard, factor or multiple standards, with or without reagent blank and/or sample blank
· Kinetics with standard, factor or multiple standards, with or without blank
· Fixed time with standard, factor or multiple standardsor with or without reagent blank · Curve fitting for non-linear standard curves
Measuring Time: · Kinetic: variable from 3 - 28 deltas, time per delta 4 - 255 sec in 1 sec steps
· Fixed time: variable from 0 - 65535 sec in 1 sec steps
Delay Time: Programmable from 0-65535 sec in 1 sec steps
Dimensions: Length 33 cm x width 34 cm x height 18 cm
Weight: 8.5 kg

 