| Type: |
Singlebeam
spectralline filterphotometer
|
|
| Light
Source: |
Tungsten
spotlight |
| Optical
Sensor: |
Silicon
photodiode |
| Wavelength: |
340
nm - 900 nm |
| Wavelength
Selection: |
Stickable
interference filters |
| Photometric
Range: |
0-2.0
Bel |
| Factor
or Standard Input: |
0.0001
to 99999 |
| Concentration:
|
0.0001
to 99999 |
| Reproducibility:
|
Short
time < 0.002 at 1 Bel / Long time < 0.010
Bel / h |
| Zero
Setting: |
Automatically
performed by key |
| Cuvette
System: |
Normal
standard cuvettes (macro or semi-micro, disposable
or special optical glass)optional: Round cuvettes
(16 mm diameter) |
| Reaction
Volume: |
500
µl per test in combination with semi-micro
cuvette 1200 µl per test in combination with
macro cuvette |
| Operator
Interface: |
Keyboard
with 20 keys for direct functions and numeric entries
|
| Data
Presentation: |
LCD
display: 2 lines of 16 characters each, easy to
read and offering simple user guidance in German
and English language |
| Memory:
|
General
operating software contained in EPROM (2 language
versions and capacity for 180 pre-programmed testvariants)
Up to 160 tests can be stored in battery-powered
RAM |
| Signal
Port: |
·
Serial port for connection to a computer (bidirectional
interface)Modes for remote control or EDP
· Connection
to external printer with serial interface
|
| Measurement
Procedures: |
·
Absorbance, Transmission
· Endpoints
with factor or standard, with or without reagent
blank and/or sample blank
· Fixed time
with factor or standard, with or without reagent
blank
· Nonlinear
calculations based on polynomials of 4th degree
|
| Measuring
Time: |
Fixed
time: variable from 5 - 300 sec in 1 sec steps |
| Delay
Time: |
Programmable
in 1 sec steps |
| Power
Supply: |
External
mains adapter: prim. 230 VAC / 50 Hz, sec. 12 VDC
/ 1 ADCDriving with 12 VDC (car-) battery is possible
|
| Dimensions:
|
Length
22 cm x width 22.8 cm x height 8.2 cm |
| Weight
Photometer: |
1.0
kg |
| Weight
Mains Adapter: |
0.9
kg |
| Certificate:
|
CE
- EC conformity markThe device complies with the
safety rules of DIN EN 60601 -1
Mains adapter with GS-sign by TÜV Rheinland
|
|
| |