| Wavelength
Range: |
190
~ 1100nm |
 |
| Spectral
Bandwidth: |
1nm |
| Wavelength
Display: |
Readable
to 0.1nm |
| Wavelength
Accuracy: |
±0.2nm
(with automatic wavelength correction) |
| Data
Bunching Interval: |
0.1,
0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9,
1.0 and 2.0nm |
| Wavelength
Reproducibility: |
±0.1nm |
| Stray
Light: |
Less
than 0.05% (at 220nm and 340nm) |
| Photometric
Range: |
Absorbance
: -0.3 ~ 4Abs |
| Photometric
System: |
Split-beam
optics, Dual detectors |
| Photometric
Accuracy: |
±0.002Abs.
at 0.5Abs. ±0.005Abs. at 1Abs. |
| Photometric
Reproducibility: |
±0.002Abs.
at 1.0Abs. |
| Baseline
Flatness: |
±0.001Abs
(2nm bandwidth, middle scanning, 1100 ~
200nm) |
|
| Baseline
Correction: |
Automatic |
| Wavelength
Slew Rate: |
About
4000nm/min |
| Light
Source: |
50W
halogen lamp (1000 hrs life) and D2 lamp
(2000 hrs life) |
| Monochromator: |
An
aberration reduced holographic concave grating
using 1200L/mm |
| Auto
Zero Function: |
By
a single stroke of AUTO ZERO key |
| Display
Graphic: |
LCD
6", adjustable viewing angle and contrast
control |
| Sample
Compartment: |
Automatic
rotative 8 position cell holder |
| Interface
Ports: |
RS-232C
and Centronics ports |
| Power
Requirements: |
85~260V~,
50/60Hz, Selected automatically |
| Dimensions: |
550(W)
X 240(H) X 400(D)mm |
| Weight: |
25kg |
MADE ACCORDING
TO CEI NORMS 66/5 |
|
|
|
|
| DESCRIPTION |
BT
600 is a conventional type uv/visible spectrophotometer
that has been satisfying many domestic and internationally
users. BT 600 is designed for user need and updated
system mostly reflects on user thought. Therefore
its mechanical or electronic upgrade will be provided
to the user in on-going basis.
|
| Precision
of Optic System |
Split
beam detection system incorporates an energy-optimized
algorithm, which gives greater photometric stability
and accuracy. |
| Easy
Operation: |
Soft
keys and also wide LCD display allow user to operate
the instrument easier than ever. |
| LCD
6" Screen: |
Via
high resolution, both test and graphic data displayed
easer performance for the repeated task. |
| Rotary
8-position cell holder: |
Auto
Controlling of 8 cells, Minimal mechanical noise,
quicker measurement time. |
| Spectrum
Scanning Mode: |
Full
Scan range from 190- 1100 nm and automatic peak/valley
determination.
To adjust the scanning interval l, 0.1, 0.2,0.5,
1.2nm.
Zoom -in function to view more accurate data.
A required graph reading is also available.
|
| DNA/Protein
Mode: |
Determine
concentration and sample purity- quickly.
To simplify your analysis, the DNA/Protein program
contains pre-configured methods for virtually all
applications. You can determine the concentration
and purity of your nucleic acid samples by selecting
either a pre-configured method or by modifying parameters
to suit your samples.
|
| Time
Course Mode: |
To
measure the changes in absorbance or transmittance
for constant wavelength with respect to time. The
changes are presented as time course curves. Besides,
obtaining real-time plots and determining; reaction
rates. Effortlessly analyze the rate reaction of
your samples. Data is plotted in real time. |
| Quantitation
Mode: |
Quickly
determine sample concentrations with linear or non
linear curve fits.
Standard curve can be formatted from 1st up to7th
linear.
BT 600 offers editing function for standard curve
Up to seven standard solvents applicable.
"Y" intercept or forced through zero can
be used
|
| PC
Control: |
PC
control s/w package via windows 95,98, 2000, NT |
| Printer
port: |
Print
out data gained from the instrument, available to
inkjet, laser printer. |
| |
|
| Continuing
research and improvements may result in specifications
change at any time. |